- Service Manual. Upgrade Document. User and Operator Manual. Customer Letter. De-Installation Manual. Installation Manual. Maintenance and Regulatory Forms.
- OEC 9800 Plus. GE Healthcare g. OEC 9800 Plus. A Classic Choice. With NewView Flat Screen Monitor. High quality images from a dependable, flexible C-arm. That's what more than. 11,000 OEC 9800 systems in use world-wide provide in thousands of life-altering procedures each. Require manual adjustments prior to shooting.
Related Manuals for GE OEC 9800 Plus. Medical Equipment GE CareScape V100 Service Manual 178 pages. Vital sings monitor. Medical Equipment GE. The OEC Elite C-arm offers superb image quality with the enhanced ease and efficiency you need for your surgical demands.
Details
Refurbishing Process:
All OEC-9800’s go through a multistep refurbishing process. Each unit is restored to a like-new status. Pain Management, Orthopedics and Rehabilitation Services are just a few of the applications utilizing the refurbished OEC-9800 c-arm. The rotating x-ray tube allows for greater penetration and resolution for the lumbar area when doing needle localizations or other work involving the thicker anatomical areas of patients. For barium swallows a VCR or Digital Video Recorder can be utilized for better record keeping and post study analysis. Igo8 software download. The 1K X 1K CCD Technology used in the OEC-9800 image system allows for crisp, clear imaging. Whatever the purpose, a refurbished OEC-9800 will deliver the results you can depend on.
All OEC-9800’s go through a multistep refurbishing process. Each unit is restored to a like-new status. Pain Management, Orthopedics and Rehabilitation Services are just a few of the applications utilizing the refurbished OEC-9800 c-arm. The rotating x-ray tube allows for greater penetration and resolution for the lumbar area when doing needle localizations or other work involving the thicker anatomical areas of patients. For barium swallows a VCR or Digital Video Recorder can be utilized for better record keeping and post study analysis. Igo8 software download. The 1K X 1K CCD Technology used in the OEC-9800 image system allows for crisp, clear imaging. Whatever the purpose, a refurbished OEC-9800 will deliver the results you can depend on.
Mechanical repairs/cosmetic maintenance
• All defective parts and covers are repaired or replaced
• C-arm is completely disassembled and repainted
• “C” is removed from main frame assembly
• Horizontal arm removed, inspected and repaired as necessary
• Vertical column removed and bearings removed inspected and replaced as required
• X-ray tube removed and inspected for artifacts and internal integrity
• Image intensifier removed and inspected for any imperfections and grid integrity
• Camera removed
• Wheels removed, cleaned and or replaced from main chassis
• Chassis inspected for structural defects and repaired as necessary
• Vertical lift assembly is tested for drift and adjusted/replaced if necessary
• All bearings are inspected, cleaned and re-lubricated, or replaced as needed
• Control panel display is tested and replaced if necessary
• All wheels are inspected and lubricated and/or replaced as needed, with new cable pushers installed
• All locks and brake assemblies are inspected, cleaned, and repaired or replaced as needed
• C-arm is completely disassembled and repainted
• Steering is tested and adjusted is required
• Foot switch and hand switch provided on each unit
• A complete set of operator and service manuals are provided
• All defective parts and covers are repaired or replaced
• C-arm is completely disassembled and repainted
• “C” is removed from main frame assembly
• Horizontal arm removed, inspected and repaired as necessary
• Vertical column removed and bearings removed inspected and replaced as required
• X-ray tube removed and inspected for artifacts and internal integrity
• Image intensifier removed and inspected for any imperfections and grid integrity
• Camera removed
• Wheels removed, cleaned and or replaced from main chassis
• Chassis inspected for structural defects and repaired as necessary
• Vertical lift assembly is tested for drift and adjusted/replaced if necessary
• All bearings are inspected, cleaned and re-lubricated, or replaced as needed
• Control panel display is tested and replaced if necessary
• All wheels are inspected and lubricated and/or replaced as needed, with new cable pushers installed
• All locks and brake assemblies are inspected, cleaned, and repaired or replaced as needed
• C-arm is completely disassembled and repainted
• Steering is tested and adjusted is required
• Foot switch and hand switch provided on each unit
• A complete set of operator and service manuals are provided
Oec 9800 Plus Service Manual Free
Re-assembly
• All components sent to C-arm reassembly area for rebuilding
• “C”, Image tube and camera, Vertical column, main control and power cable installed
• New batteries are installed
• New CRT monitors installed in monitor cart
• Monitor cart rebuilt
• When C-arm is completely rebuilt mechanically system is sent for calibration and final testing
• All components sent to C-arm reassembly area for rebuilding
• “C”, Image tube and camera, Vertical column, main control and power cable installed
• New batteries are installed
• New CRT monitors installed in monitor cart
• Monitor cart rebuilt
• When C-arm is completely rebuilt mechanically system is sent for calibration and final testing
Sketchup pro 2019 19 1. Testing/calibration
• Image intensifier and X-ray tube are tested for stability and balance
• Image intensifier is tested for resolution and gain to be within OEM specifications
• X-ray tube bearings are tested for noise and coast time
• X-ray tube filaments and stator windings are tested
• X-ray tube radiation output is verified to be within OEM specifications
• X-ray generator high voltage tested
• Maximum dose rate is set in compliance with government regulations
• CCD camera calibrated to OEM specifications
• X-ray beam is aligned for each field size
• C-arm is tested for current leakage requirements
• All monitors are aligned, resized, centered and focused
• All monitors are adjusted for brightness and linearity
• Video system is optimized for gray scale and resolution
• All power supplies are tested and calibrated or replaced as necessary
• Image intensifier and X-ray tube are tested for stability and balance
• Image intensifier is tested for resolution and gain to be within OEM specifications
• X-ray tube bearings are tested for noise and coast time
• X-ray tube filaments and stator windings are tested
• X-ray tube radiation output is verified to be within OEM specifications
• X-ray generator high voltage tested
• Maximum dose rate is set in compliance with government regulations
• CCD camera calibrated to OEM specifications
• X-ray beam is aligned for each field size
• C-arm is tested for current leakage requirements
• All monitors are aligned, resized, centered and focused
• All monitors are adjusted for brightness and linearity
• Video system is optimized for gray scale and resolution
• All power supplies are tested and calibrated or replaced as necessary
Additional Information
Oec 9800 Manual
Specifications | Terraria 1.3 4.4 mac download. Technical Specificaitons Generator • High frequency generator 15kW full-wave • Up to 120 KVP and 75MA for radiographic exposures • Continuous fluro-mode up to 20 mA • One shot frame integration (low, medium, high) • Full power from standard 110V 15A Or 220V 8A outlet • Patented energy buffer design X-Ray Tube • Rotating anode X-ray tube • Focal spots: 0.3 - 0.6 mm • 300,000 HU Anode heat capapcity Express scribe 5.63 keygen. Physical Specificaitons • Free Space in Arc - 31” • Depth of Arc: 26” • Arc orbital movement: 115 Degrees • Left/Right wig-wag scan: +/- 20 degrees • Vertical Travel: 18” motorized • Horizontal travel: 8” • Lateral rotation: +/- 180 degrees |
---|---|
File Attachments | No |